Monte Curcio instruments. API 700 - Mass flow calibrator
Monte Curcio (MCU) instruments: API Teledyne 700 mass flow calibrator. The Model M700 can generate diluted calibration gas including ozone, gas phase titration (NO2) from gas cylinders, or from a permeation tube source gas. Each concentration is generated precisely by mixing the gas with diluent zero air in which the mixing ratio is controlled by the mass flow controller. The diluent air mass flow controller has a range of 10 SLPM (20 SLPM option), and the gas mass flow controller has a range of 100 cc/min. (0-50 cc/min., 0-200cc/min. options). The mass flow controllers assure a precise mixing ratio for accurate and precise calibration gas generation using the state-of- the-art electronic closed-loop control.
|
Citation proposal
Pirrone N. (CNR-IIA, Institute of Atmospheric Pollution Research). Monte Curcio instruments. API 700 - Mass flow calibrator. CNR-Institute of Atmospheric Pollution Research https://geoportal.mountaingenius.org/geonetwork/srv/api/records/mcu_inst_api_700_cal |
Simple
- Date ( Creation )
- 2015-02-25
- Edition
- First
- Identifier
- mt_curcio / mcu_inst_api_700_cal
- Purpose
- Assessment of environmental pollution
Principal investigator
- Maintenance and update frequency
- Not planned
- GEOSS Societal Benefit Areas, version 2.3
-
- Health
- GEMET - INSPIRE themes, version 1.0
-
- Environmental monitoring facilities , Human health and safety
- GEMET - Concepts, version 2.4
-
- pollution , pollutant
- GEMET - Groups, version 2.4
-
- CHEMISTRY, SUBSTANCES, PROCESSES
- GEOSS - Earth Observation Vocabulary, version 1.0
-
- ATMOSPHERIC CHEMISTRY , Contaminants/Pollutants (Inorganic/Organic)
- GEOSS - Societal Benefit Areas, version 1.0
-
- Global biogeochemistry , Human health , Pollution events
- EARTh - Environmental Applications Reference Thesaurus
-
- monitoring stations , pollution monitoring , environmental monitoring , monitoring network , air quality monitoring , pollutant monitoring
- Use limitation
- geossNoMonetaryCharge
- Access constraints
- Other restrictions
- Other constraints
- (e) intellectual property rights
- Spatial representation type
- Vector
- Denominator
- 1000
- Metadata language
- English
- Character set
- UTF8
- Topic category
-
- Environment
- Health
N
S
E
W
))
- Time period
- 2015-01-012015-01-01
- Reference system identifier
-
EPSG 4326
- Date ( Revision )
- 2009-10-15
- Distribution format
-
- unknown (unknown )
- OnLine resource
- Browser to MCU databases
- OnLine resource
- http://sdi.iia.cnr.it/mcugeoserver/ows?
- OnLine resource
- http://sdi.iia.cnr.it/gmosgeoserver/ows?
- Hierarchy level
- Dataset
Conformance result
- Date ( Publication )
- 2009-06-11
- Explanation
- See the referenced specification
- Pass
- true
- Statement
- Maintenance requires visually inspection to verify that the system is in order. Options such as the ozone generator, permeation oven, or the photometer flows are controlled by the critical flow orifices. The orifices never need adjustment. The critical flow orifice maintains precise volumetric control as long as the critical pressure is maintained between upstream and downstream of the orifice. The sample flow across the internal pneumatic system is fixed by the critical flow control orifice and has no adjustment. The following procedures are identical for each type of orifice assembly.//Ref. Operation manual Model 700 (http://www.teledyne-api.com/manuals/)
gmd:MD_Metadata
- File identifier
- mcu_inst_api_700_cal XML
- Metadata language
- English
- Character set
- UTF8
- Parent identifier
- b123e1e2-95d7-46c1-a476-b2de45c8fc33
- Hierarchy level
- Collection hardware
- Date stamp
- 2015-06-19T17:26:28
- Metadata standard name
- ISO 19115:2003/19139
- Metadata standard version
- 1.0
Overviews

thumbnail

large_thumbnail
Provided by

Views
mcu_inst_api_700_cal
Access to the portal
Read here the full details and access to the data.
Associated resources
Not available