• SHARE GeoNetwork
  •  
  •    

  Monte Curcio instruments. TSI Scanning Mobility Particle Sizer

Monte Curcio (MCU) instruments: TSI Scanning Mobility Particle Sizer (SMPS). The Model SMPS is designed for long-term measurement of aerosol particles and over there in highly burden regions. It is a mobility particle size spectrometer, which can measure continuously the particle number size distribution in a size range from 10 to 800 Nanometres. The temporal resolution to measure a complete particle number size distribution is approximately 5 minutes. Equipment is based on TSI CPC 3775.
 
Citation proposal
Pirrone N. (CNR-IIA, Institute of Atmospheric Pollution Research). Monte Curcio instruments. TSI Scanning Mobility Particle Sizer. CNR-Institute of Atmospheric Pollution Research https://geoportal.mountaingenius.org/geonetwork/srv/api/records/mcu_inst_TSI_SMPS_pm
 

Simple

Date ( Creation )
2015-02-25
Edition
First
Identifier
mt_curcio / mcu_inst_TSI_SMPS_pm
Purpose
Assessment of environmental pollution

  Principal investigator

CNR-IIA, Institute of Atmospheric Pollution Research - Sprovieri F.  

  Custodian

CNR-IIA, Institute of Atmospheric Pollution Research - Pirrone N.  

  Publisher

CNR-Institute of Atmospheric Pollution Research - Pirrone N.  

Maintenance and update frequency
Not planned
GEOSS Societal Benefit Areas, version 2.3
  • Health
GEMET - INSPIRE themes, version 1.0
  • Environmental monitoring facilities , Human health and safety
GEMET - Concepts, version 2.4
  • pollution , pollutant
GEMET - Groups, version 2.4
  • CHEMISTRY, SUBSTANCES, PROCESSES
GEOSS - Earth Observation Vocabulary, version 1.0
  • ATMOSPHERIC CHEMISTRY , Contaminants/Pollutants (Inorganic/Organic)
GEOSS - Societal Benefit Areas, version 1.0
  • Global biogeochemistry , Human health , Pollution events
EARTh - Environmental Applications Reference Thesaurus
  • monitoring stations , pollution monitoring , environmental monitoring , monitoring network , air quality monitoring , pollutant monitoring
Use limitation
geossNoMonetaryCharge
Access constraints
Other restrictions
Other constraints
(e) intellectual property rights
Spatial representation type
Vector
Denominator
1000
Metadata language
English
Character set
UTF8
Topic category
  • Environment
  • Health
N
S
E
W


Time period
2015-01-012015-01-01
Reference system identifier
EPSG 4326

 

Date ( Revision )
2009-10-15

Distribution format
  • unknown (unknown )

OnLine resource
Browser to MCU databases  
OnLine resource
http://sdi.iia.cnr.it/mcugeoserver/ows?  
OnLine resource
http://sdi.iia.cnr.it/gmosgeoserver/ows?  
Hierarchy level
Dataset

Conformance result

Date ( Publication )
2009-06-11
Explanation
See the referenced specification
Pass
true
Statement
Annual calibration of the instrument is performed by sizing accuracy with size standard and calibration of the sheath air flow rate. A size standard is used to determine the exact sizing of the instrument and therefore to calibrate the sheath air flow rate. This procedure is performed using an ultrasonic nebulizer, pure water, and a 200 nm size standard. Calibration of the pressure transducer is performed annually as well as check of the saturator sponge inside the CPC.//Ref. User manual of TROPOS-SMPS (http://wiki.tropos.de/index.php/SMPS_-_DMPS_%28Mobility_Particle_Size_Spectrometer%29)

gmd:MD_Metadata

File identifier
mcu_inst_TSI_SMPS_pm   XML
Metadata language
English
Character set
UTF8
Parent identifier
b123e1e2-95d7-46c1-a476-b2de45c8fc33
Hierarchy level
Collection hardware
Date stamp
2015-06-19T17:43:45
Metadata standard name
ISO 19115:2003/19139
Metadata standard version
1.0

  Originator

CNR-Institute of Atmospheric Pollution Research - Cinnirella S.  

 
 

  Overviews

thumbnail
large_thumbnail

  Provided by

  Share on social sites

         

  Views

  • INSPIRE
  • Simple
  • Full
  • XML
mcu_inst_TSI_SMPS_pm   Access to the portal Read here the full details and access to the data.