Monte Curcio instruments. TSI Scanning Mobility Particle Sizer
Monte Curcio (MCU) instruments: TSI Scanning Mobility Particle Sizer (SMPS). The Model SMPS is designed for long-term measurement of aerosol particles and over there in highly burden regions. It is a mobility particle size spectrometer, which can measure continuously the particle number size distribution in a size range from 10 to 800 Nanometres. The temporal resolution to measure a complete particle number size distribution is approximately 5 minutes. Equipment is based on TSI CPC 3775.
|
Citation proposal
Pirrone N. (CNR-IIA, Institute of Atmospheric Pollution Research). Monte Curcio instruments. TSI Scanning Mobility Particle Sizer. CNR-Institute of Atmospheric Pollution Research https://geoportal.mountaingenius.org/geonetwork/srv/api/records/mcu_inst_TSI_SMPS_pm |
Simple
- Date ( Creation )
- 2015-02-25
- Edition
- First
- Identifier
- mt_curcio / mcu_inst_TSI_SMPS_pm
- Purpose
- Assessment of environmental pollution
Principal investigator
- Maintenance and update frequency
- Not planned
- GEOSS Societal Benefit Areas, version 2.3
-
- Health
- GEMET - INSPIRE themes, version 1.0
-
- Environmental monitoring facilities , Human health and safety
- GEMET - Concepts, version 2.4
-
- pollution , pollutant
- GEMET - Groups, version 2.4
-
- CHEMISTRY, SUBSTANCES, PROCESSES
- GEOSS - Earth Observation Vocabulary, version 1.0
-
- ATMOSPHERIC CHEMISTRY , Contaminants/Pollutants (Inorganic/Organic)
- GEOSS - Societal Benefit Areas, version 1.0
-
- Global biogeochemistry , Human health , Pollution events
- EARTh - Environmental Applications Reference Thesaurus
-
- monitoring stations , pollution monitoring , environmental monitoring , monitoring network , air quality monitoring , pollutant monitoring
- Use limitation
- geossNoMonetaryCharge
- Access constraints
- Other restrictions
- Other constraints
- (e) intellectual property rights
- Spatial representation type
- Vector
- Denominator
- 1000
- Metadata language
- English
- Character set
- UTF8
- Topic category
-
- Environment
- Health
N
S
E
W
))
- Time period
- 2015-01-012015-01-01
- Reference system identifier
-
EPSG 4326
- Date ( Revision )
- 2009-10-15
- Distribution format
-
- unknown (unknown )
- OnLine resource
- Browser to MCU databases
- OnLine resource
- http://sdi.iia.cnr.it/mcugeoserver/ows?
- OnLine resource
- http://sdi.iia.cnr.it/gmosgeoserver/ows?
- Hierarchy level
- Dataset
Conformance result
- Date ( Publication )
- 2009-06-11
- Explanation
- See the referenced specification
- Pass
- true
- Statement
- Annual calibration of the instrument is performed by sizing accuracy with size standard and calibration of the sheath air flow rate. A size standard is used to determine the exact sizing of the instrument and therefore to calibrate the sheath air flow rate. This procedure is performed using an ultrasonic nebulizer, pure water, and a 200 nm size standard. Calibration of the pressure transducer is performed annually as well as check of the saturator sponge inside the CPC.//Ref. User manual of TROPOS-SMPS (http://wiki.tropos.de/index.php/SMPS_-_DMPS_%28Mobility_Particle_Size_Spectrometer%29)
gmd:MD_Metadata
- File identifier
- mcu_inst_TSI_SMPS_pm XML
- Metadata language
- English
- Character set
- UTF8
- Parent identifier
- b123e1e2-95d7-46c1-a476-b2de45c8fc33
- Hierarchy level
- Collection hardware
- Date stamp
- 2015-06-19T17:43:45
- Metadata standard name
- ISO 19115:2003/19139
- Metadata standard version
- 1.0
Overviews

thumbnail

large_thumbnail
Provided by

Views
mcu_inst_TSI_SMPS_pm
Access to the portal
Read here the full details and access to the data.
Associated resources
Not available