• SHARE GeoNetwork
  •  
  •    

  Pyramid air temperature

At the Pyramid Laboratory Observatory AWS the DMA570 Lsi-Lastem Thermohygrometer is mounted on a 2 m pole.

It is a probe for measuring air temperature and air relative humidity with replaceable sensitive element for simplify calibration and maintenance.

 
Citation proposal
(2011) . Pyramid air temperature. https://geoportal.mountaingenius.org/geonetwork/srv/api/records/71a5f1cb-e9e0-44f1-8e5d-aa1adc8ed49c
 

Simple

Date ( Publication )
2011-03-30T17:51:00
Purpose

The purpose of this resource is to acquire data on:

- air temperature

- relative humidity

Status
Completed

  Point of contact

Ev-K2-CNR  
Via San Bernardino 145 Bergamo BG 24126 Italy

  +39 035 3230511  
  +39 035 3230551 
Website
http://www.evk2cnr.org/  
Maintenance and update frequency
As needed
GEMET ( Theme )
  • thermal power plant
Keywords
  • SHARE , Network , Station , High Altitude
Keywords ( Place )
  • Nepal , Lobuche , Khumbu Valley , Sagarmatha National Park
Keywords ( Theme )
  • AWS
Keywords ( Theme )
  • air temperature , relative humidity
Use limitation
No conditions apply
Access constraints
Copyright
Other constraints
Character set
UTF8
Topic category
  • Climatology, meteorology, atmosphere
  • Environment
  • Geoscientific information
N
S
E
W


Supplemental Information
Reference system identifier
EPSG / WGS 84 (EPSG:4326) / 7.4
Topology level
Geometry only
Geometric object type
Point
Geometric object count
1
OnLine resource
Himalayas Reference Site Pyramid Station  

Himalayas Reference Site Pyramid Station

OnLine resource
https://geoportal.mountaingenius.org/portal/index.php/data/  
OnLine resource
https://geoportal.mountaingenius.org/geonetwork/srv/api/records/71a5f1cb-e9e0-44f1-8e5d-aa1adc8ed49c/attachments/Data%20Usage%20Policy.pdf  
Hierarchy level
Series
Statement

At the Pyramid Laboratory Observatory AWS the DMA570 Lsi-Lastem Thermohygrometer is mounted on a 2 m pole.

It is a probe for measuring air temperature and air relative humidity with replaceable sensitive element for simplify calibration and maintenance. LSI-Lastem supplies a precise and reliable set of probes, suitable for a continuous measurement in severe environments, in presence of deep thermal and hygrometric ranges with high sun radiant heat. An important feature of this set of sensor is that the thermohygrometric sensitive element is easily replaceable, in order to have a simple and rapid ordinary maintenance and avoiding calibration. The Thermohygrometer has supported a considerable improvement, owing to a deep technical and styling development process: a fan ensures a continuos air change around the sensor in order to eliminate temperature fault caused by radiant heat.

For temperature:

Range -30.+70°C

Sensitive element Pt100 1/3 DIN-B

Accuracy ± 0,1°C (0°C)

Hysteresis&Repeatability na

Resolution 0,025°C

Long term stability -0.04% (after 5 years at 200°C)

Calibration uncertainty 0,1°C

Response time (T63) Wind speed 0,3-0,5 m/s: 80sec

Radiation shield efficiency with respect to forced ventilation sensor (wind 0,5m/s, rad.800 W/m2): +0,6°C

For relative humidity:

Range: Nom.0..100%, Eff.10..98%

Sensitive element Capacitive

Accuracy 2,5% (11-90%) 0,6°C (20°C, 50%RH)

Hysteresis&Repeatability 0,5%

Resolution 0,2%

Long term stability -2%/year (at 75%RH) n.a.

Calibration uncertainty 0,1°C 1,5% na

Response time (T63) Wind speed 0,3-0,5 m/s: 80sec

Radiation shield efficiency with respect to forced ventilation sensor (wind 0,5m/s, rad.800 W/m2): n.a.

General characteristic:

Ventilation: Natural

Electric output Jumper locally selectable 2 x 0-20 mA, 4-20 mA, 0-5 V, 1-5 V. 60-300 mV

Output signals Output n.1: temperature. Output n.2: RH% or dew point (default RH%)

Operating temperature -30°+70°C

Sampling rate 1 sec. (default) PC programmable 1..300 sec.

N°acquisitions for mobile average calculation: N°1 (default), PC programmable 1..20 acquisitions

Load resistance (mA output): DMA570 300 Ohm

gmd:MD_Metadata

File identifier
71a5f1cb-e9e0-44f1-8e5d-aa1adc8ed49c   XML
Metadata language
eng
Character set
UTF8
Parent identifier
d156500b-6e83-40e9-abb1-a87e53ea0224
Hierarchy level
Series
Date stamp
2025-02-07T16:36:07
Metadata standard name
ISO 19115:2003/19139
Metadata standard version
1.0

  Point of contact

Ev-K2-CNR  
Via San Bernardino 145 Bergamo BG 24126 Italy

  +39 035 3230511  
  +39 035 3230551 
Website
http://www.evk2cnr.org  
 
 

  Overviews

evk2cnr__.png
evk2cnr__.png

  Provided by

  Share on social sites

         

  Views

  • INSPIRE
  • Simple
  • Full
  • XML
71a5f1cb-e9e0-44f1-8e5d-aa1adc8ed49c   Access to the portal Read here the full details and access to the data.

  Associated resources

Not available


  •  
  •  
  • API