  Namche Thermohygrometer

At the Namche AWS the DMA570 Lsi-Lastem Thermohygrometer is mounted on a 2 m pole.

It is a probe for measuring air temperature and air relative humidity with replaceable sensitive element for simplify calibration and maintenance. LSI-Lastem supplies a precise and reliable set of probes, suitable for a continuous measurement in severe environments, in presence of deep thermal and hygrometric ranges with high sun radiant heat. An important feature of this set of sensor is that the thermohygrometric sensitive element is easily replaceable, in order to have a simple and rapid ordinary maintenance and avoiding calibration. The Thermohygrometer has supported a considerable improvement, owing to a deep technical and styling development process: a fan ensures a continuos air change around the sensor in order to eliminate temperature fault caused by radiant heat.

For temperature:

Range -30.+70°C

Sensitive element Pt100 1/3 DIN-B

Accuracy ± 0,1°C (0°C)

Hysteresis&Repeatability na

Resolution 0,025°C

Long term stability -0.04% (after 5 years at 200°C)

Calibration uncertainty 0,1°C

Response time (T63) Wind speed 0,3-0,5 m/s: 80sec

Radiation shield efficiency with respect to forced ventilation sensor (wind 0,5m/s, rad.800 W/m2): +0,6°C

For relative humidity:

Range: Nom.0..100%, Eff.10..98%

Sensitive element Capacitive

Accuracy 2,5% (11-90%) 0,6°C (20°C, 50%RH)

Hysteresis&Repeatability 0,5%

Resolution 0,2%

Long term stability -2%/year (at 75%RH) n.a.

Calibration uncertainty 0,1°C 1,5% na

Response time (T63) Wind speed 0,3-0,5 m/s: 80sec

Radiation shield efficiency with respect to forced ventilation sensor (wind 0,5m/s, rad.800 W/m2): n.a.

General characteristic:

Ventilation: Natural

Electric output Jumper locally selectable 2 x 0-20 mA, 4-20 mA, 0-5 V, 1-5 V. 60-300 mV

Output signals Output n.1: temperature. Output n.2: RH% or dew point (default RH%)

Operating temperature -30°+70°C

Sampling rate 1 sec. (default) PC programmable 1..300 sec.

N°acquisitions for mobile average calculation: N°1 (default), PC programmable 1..20 acquisitions

Load resistance (mA output): DMA570 300 Ohm

 
Citation proposal
(2011) . Namche Thermohygrometer. https://geoportal.mountaingenius.org/geonetwork/srv/api/records/5f448357-08b6-4d45-b57a-4fc36f721d02
 

Simple

Date ( Publication )
2011-05-20T03:00:00
Purpose

Meteo-climatic observations in this area are very important for the study of the environmental changes that are affecting this ecosystem, helping the scientific community to improve the prediction of the evolution of this environmental phenomena and modification.

This station is reference point within the CEOP-High Elevation (CEOP-HE) initiative, promoted in order to investigate possible changes of hydrological and energy cycles in high elevation regions and to better elucidate the role of mountains within the climate system from a local to global scales.

The purpose of this resource in Namche station is to acquire data on:

- air temperature

- relative humidity

Status
Completed

  Point of contact

Ev-K2-CNR  
Via San Bernardino 145 Bergamo BG 24126 Italy

  +39 035 3230511  
  +39 035 3230551 
Website
http://www.evk2cnr.org/  
Maintenance and update frequency
As needed
GEMET ( Theme )
  • thermal power plant
Keywords
  • SHARE , Monitoring Network , Station , High Altitude
Keywords
  • CEOP
Keywords ( Place )
  • Namche , Khumbu Valley , Sagarmatha National Park , Nepal
Keywords ( Theme )
  • air temperature
Use limitation
No conditions apply
Access constraints
Copyright
Other constraints
Metadata language
eng
Character set
UTF8
Topic category
  • Climatology, meteorology, atmosphere
  • Environment
  • Geoscientific information
N
S
E
W


Supplemental Information
Reference system identifier
EPSG / WGS 84 (EPSG:4326) / 7.4
Topology level
Geometry only
Geometric object type
Point
Geometric object count
1
OnLine resource
Himalayas Reference Site Namche Station  

Himalayas Reference Site Namche Station

OnLine resource
https://geoportal.mountaingenius.org/geonetwork/srv/api/records/5f448357-08b6-4d45-b57a-4fc36f721d02/attachments/Data%20Usage%20Policy.pdf  
Hierarchy level
Series
Other
Instruments: AWS sensors
Statement

The station was installed on October 2001. The sensors are mounted on a 2m masts

For temperature:

Range -30.+70°C

Sensitive element Pt100 1/3 DIN-B

Accuracy ± 0,1°C (0°C)

Hysteresis&Repeatability na

Resolution 0,025°C

Long term stability -0.04% (after 5 years at 200°C)

Calibration uncertainty 0,1°C

Response time (T63) Wind speed 0,3-0,5 m/s: 80sec

Radiation shield efficiency with respect to forced ventilation sensor (wind 0,5m/s, rad.800 W/m2): +0,6°C

For relative humidity:

Range: Nom.0..100%, Eff.10..98%

Sensitive element Capacitive

Accuracy 2,5% (11-90%) 0,6°C (20°C, 50%RH)

Hysteresis&Repeatability 0,5%

Resolution 0,2%

Long term stability -2%/year (at 75%RH) n.a.

Calibration uncertainty 0,1°C 1,5% na

Response time (T63) Wind speed 0,3-0,5 m/s: 80sec

Radiation shield efficiency with respect to forced ventilation sensor (wind 0,5m/s, rad.800 W/m2): n.a.

General characteristic:

Ventilation: Natural

Electric output Jumper locally selectable 2 x 0-20 mA, 4-20 mA, 0-5 V, 1-5 V. 60-300 mV

Output signals Output n.1: temperature. Output n.2: RH% or dew point (default RH%)

Operating temperature -30°+70°C

Sampling rate 1 sec. (default) PC programmable 1..300 sec.

N°acquisitions for mobile average calculation: N°1 (default), PC programmable 1..20 acquisitions

Load resistance (mA output): DMA570 300 Ohm

gmd:MD_Metadata

File identifier
5f448357-08b6-4d45-b57a-4fc36f721d02   XML
Metadata language
eng
Character set
UTF8
Parent identifier
2e8cbdb5-bf79-40f2-86bc-d26050336ff5
Hierarchy level
Series
Date stamp
2025-02-24T08:52:44
Metadata standard name
ISO 19115:2003/19139
Metadata standard version
1.0

  Point of contact

Ev-K2-CNR  
Via San Bernardino 145 Bergamo BG 24126 Italy

  +39 035 3230511  
  +39 035 3230551 
Website
http://www.evk2cnr.org  
 
 

  Overviews

evk2cnr__.png

  Provided by

  Share on social sites

         

  Views

  • INSPIRE
  • Simple
  • Full
  • XML
5f448357-08b6-4d45-b57a-4fc36f721d02   Access to the portal Read here the full details and access to the data.

  Associated resources

Not available


  •  
  •  
  • API