Forni Thermohygrometer
At the Forni AWS the DMA570 Lsi-Lastem Thermohygrometer is mounted on a 2 m pole.
It is a probe for measuring air temperature and air relative humidity with replaceable sensitive element for simplify calibration and maintenance. LSI-Lastem supplies a precise and reliable set of probes, suitable for a continuous measurement in severe environments, in presence of deep thermal and hygrometric ranges with high sun radiant heat. An important feature of this set of sensor is that the thermohygrometric sensitive element is easily replaceable, in order to have a simple and rapid ordinary maintenance and avoiding calibration. The Thermohygrometer has supported a considerable improvement, owing to a deep technical and styling development process: a fan ensures a continuos air change around the sensor in order to eliminate temperature fault caused by radiant heat.
For temperature:
Range -30.+70°C
Sensitive element Pt100 1/3 DIN-B
Accuracy ± 0,1°C (0°C)
Hysteresis&Repeatability na
Resolution 0,025°C
Long term stability -0.04% (after 5 years at 200°C)
Calibration uncertainty 0,1°C
Response time (T63) Wind speed 0,3-0,5 m/s: 80sec
Radiation shield efficiency with respect to forced ventilation sensor (wind 0,5m/s, rad.800 W/m2): +0,6°C
For relative humidity:
Range: Nom.0..100%, Eff.10..98%
Sensitive element Capacitive
Accuracy 2,5% (11-90%) 0,6°C (20°C, 50%RH)
Hysteresis&Repeatability 0,5%
Resolution 0,2%
Long term stability -2%/year (at 75%RH) n.a.
Calibration uncertainty 0,1°C 1,5% na
Response time (T63) Wind speed 0,3-0,5 m/s: 80sec
Radiation shield efficiency with respect to forced ventilation sensor (wind 0,5m/s, rad.800 W/m2): n.a.
General characteristic:
Ventilation: Natural
Electric output Jumper locally selectable 2 x 0-20 mA, 4-20 mA, 0-5 V, 1-5 V. 60-300 mV
Output signals Output n.1: temperature. Output n.2: RH% or dew point (default RH%)
Operating temperature -30°+70°C
Sampling rate 1 sec. (default) PC programmable 1..300 sec.
N°acquisitions for mobile average calculation: N°1 (default), PC programmable 1..20 acquisitions
Load resistance (mA output): DMA570 300 Ohm
|
Citation proposal
(2011) . Forni Thermohygrometer. https://geoportal.mountaingenius.org/geonetwork/srv/api/records/55638813-8255-4128-bb24-d4a246d45910 |
INSPIRE
Identification
- File identifier
- 55638813-8255-4128-bb24-d4a246d45910 XML
- Hierarchy level
- Series
Online resource
- Protocol
- OGC:WMS-1.1.1-http-get-map
- Protocol
- GLG:KML-2.0-http-get-map
Resource identifier
- Spatial representation type
- Vector
Encoding
Projection
- code
- WGS 84 (EPSG:4326)
Classification of data and services
- Topic category
-
- Climatology, meteorology, atmosphere
- Geoscientific information
- Environment
Classification of data and services
Coupled resource
Coupled resource
Classification of data and services
Coupled resource
Coupled resource
Keywords
- GEMET themes ( Theme )
-
- air temperature, relative humidity
- GEMET themes ( Theme )
-
- air temperature, relative humidity
- GEMET themes ( Theme )
-
- air temperature, relative humidity
- GEMET themes ( Theme )
-
- air temperature, relative humidity
Other keywords
- Keywords ( Theme )
-
- NEXTDATA
- Keywords ( Theme )
-
- NEXTDATA
- Keywords ( Theme )
-
- Keywords ( Theme )
-
- Keywords ( Theme )
-
- SHARE , network
- Keywords ( Place )
-
- Forni Glacier, Ortles-Cevedale Group, Stelvio National Parks, Alps, Italy
- Keywords ( Theme )
-
- NEXTDATA
- Keywords ( Theme )
-
- Keywords ( Theme )
-
- Keywords ( Theme )
-
- SHARE , network
- Keywords ( Place )
-
- Forni Glacier, Ortles-Cevedale Group, Stelvio National Parks, Alps, Italy
- Keywords ( Theme )
-
- NEXTDATA
- Keywords ( Theme )
-
- Keywords ( Theme )
-
- Keywords ( Theme )
-
- SHARE , network
- Keywords ( Place )
-
- Forni Glacier, Ortles-Cevedale Group, Stelvio National Parks, Alps, Italy
- Keywords ( Theme )
-
- NEXTDATA
- Keywords ( Theme )
-
- Keywords ( Theme )
-
- Keywords ( Theme )
-
- SHARE , network
- Keywords ( Place )
-
- Forni Glacier, Ortles-Cevedale Group, Stelvio National Parks, Alps, Italy
Geographic coverage
))
Temporal reference
Temporal extent
Temporal extent
- Date ( Publication )
- 2011-04-12T19:08:00
Quality and validity
- Lineage
- This sensor is mounted on a mast at 2 m of height.
Conformity
Conformity
Conformity
Conformity
Conformity
Conformity
Restrictions on access and use
- Access constraints
Restrictions on access and use
Responsible organization (s)
Contact for the resource
- Organisation name
- Department of Earth Science "Ardito Desio"
Responsible organization (s)
Contact for the resource
- Organisation name
- Department of Earth Science "Ardito Desio"
Metadata information
Contact for the metadata
- Organisation name
- Ev-K2-CNR
- metadata@evk2cnr.org
- Date stamp
- 2025-02-24T09:17:21
- Metadata language
- eng
- Character set
- UTF8
SDS
Conformance class 1: invocable
Access Point URL
Endpoint URL
Technical specification
Conformance class 2: interoperable
Coordinate reference system
Quality of Service
Access constraints
Limitation
Use constraints
Limitation
Responsible custodian
Contact for the resource
Conformance class 3: harmonized
Overviews
Provided by