• SHARE GeoNetwork
  •  
  •    

  Gas Chromatograph RGD

A Gas-Chromatograph equipped with an RGD detector to measure carbon monoxide (CO).

Measurements started on Jan 2007 with a modified version of a commercial RGD2-Trace Analytical analyser. Measurements ended in March 2010.

 
Citation proposal
(2011) . Gas Chromatograph RGD. https://geoportal.mountaingenius.org/geonetwork/srv/api/records/3fe7953d-daea-4a62-becd-6787f96f4405
 
  • INSPIRE
  • SDS

INSPIRE

Identification

File identifier
3fe7953d-daea-4a62-becd-6787f96f4405   XML
Hierarchy level
Series
Online resource
Protocol
WWW:LINK-1.0-http--link
Protocol
OGC:WMS-1.1.1-http-get-map
Protocol
GLG:KML-2.0-http-get-map
Protocol
WWW:DOWNLOAD-1.0-http--download
Protocol
WWW:DOWNLOAD-1.0-http--download
Resource identifier
Spatial representation type
Vector
Encoding
Projection
code
WGS 1984
 

Classification of data and services

Topic category
  • Climatology, meteorology, atmosphere
  • Geoscientific information
  • Environment
 

Classification of data and services

Coupled resource
Coupled resource
 
 

Classification of data and services

Coupled resource
Coupled resource
 
 

Keywords

Other keywords
Keywords ( Theme )
  • NEXTDATA
Keywords ( Theme )
  • NEXTDATA
Keywords ( Theme )
Keywords
  • Carbon monoxide , Gas-cromatography , Station , High Altitude , SHARE , Network , Apennines
Keywords
  • GAW , Atmospheric conditionsEE , Environmental monitoring facilities
Keywords
  • Monte Cimone
Keywords ( Theme )
  • NEXTDATA
Keywords ( Theme )
Keywords
  • Carbon monoxide , Gas-cromatography , Station , High Altitude , SHARE , Network , Apennines
Keywords
  • GAW , Atmospheric conditionsEE , Environmental monitoring facilities
Keywords
  • Monte Cimone
Keywords ( Theme )
  • NEXTDATA
Keywords ( Theme )
Keywords
  • Carbon monoxide , Gas-cromatography , Station , High Altitude , SHARE , Network , Apennines
Keywords
  • GAW , Atmospheric conditionsEE , Environmental monitoring facilities
Keywords
  • Monte Cimone
Keywords ( Theme )
  • NEXTDATA
Keywords ( Theme )
Keywords
  • Carbon monoxide , Gas-cromatography , Station , High Altitude , SHARE , Network , Apennines
Keywords
  • GAW , Atmospheric conditionsEE , Environmental monitoring facilities
Keywords
  • Monte Cimone
 
 

Geographic coverage

N
S
E
W


 

Temporal reference

Temporal extent
Temporal extent
Date ( Publication )
2011-04-05T03:00:00
 

Quality and validity

Lineage

The instrument is a gas-chromatograph equipped with an RGD detector.

The instrument is running continuously.

In Sampling and Analysis Frequency the high flux of real air continuosly flow through the sampling line and every 30 minutes a portion is splitted for 3 minutes to the gas-chromatograph sampling loop.

Description of Instruments:

time resolution: 1 measurement every 30 minutes;

sensitivity: +- 1 ppb

detection limit: 10 ppb

measurements range: 1 - 10^3 (10 ppb - 10 ppm)

Time zone Local time UTC+1

Processing/Calibration:

-Current Scale: indirect link to NOAA 2004 scale

-Scale and Calibration (traceability): reference standard are real air pumped in Luxfer Scott Marrin cylinders and calibrated at MPI-Jena against NOAA cylinders.

-Measurement Calibration: every 15 minutes a sample is analyzed alternating between real air and reference standard in order to correct for short time instrumental drifts.

Molar fraction is calculated from the ratio of the peak heights and than corrected for non linearity of the detector.

 

Conformity

Conformity
Conformity
 

Conformity

Conformity
Conformity
 

Restrictions on access and use

Access constraints
 

Restrictions on access and use

 

Responsible organization (s)

Contact for the resource
Organisation name
Urbino University Dipartimento di Scienza di Base e Fondamenti
Email
michela.maione@uniurb.it
Organisation name
Urbino University Dipartimento di Scienza di Base e Fondament
Email
jgor.arduini@uniurb.it
 

Responsible organization (s)

Contact for the resource
Organisation name
Urbino University Dipartimento di Scienza di Base e Fondamenti
Email
michela.maione@uniurb.it
Organisation name
Urbino University Dipartimento di Scienza di Base e Fondament
Email
jgor.arduini@uniurb.it
 

Metadata information

Contact for the metadata
Organisation name
National Research Council (CNR – Consiglio Nazionale delle Ricerche). Institute for Atmospheric Science and Climate (ISAC – Istituto di Scienze dell’Atmosfera e del Clima)
Email
p.bonasoni@isac.cnr.it
Organisation name
Official website of the Mount Cimone Station - Italian Climate Observatory "Ottavio Vittori"
Email
michela.maione@uniurb.it
Date stamp
2019-12-04T19:15:35
Metadata language
eng
Character set
UTF8
 
 

SDS

Conformance class 1: invocable

Access Point URL
Endpoint URL
Technical specification
 

Conformance class 2: interoperable

Coordinate reference system
 
Quality of Service
 
Access constraints
Limitation
 
Use constraints
Limitation
 
Responsible custodian
Contact for the resource
 
 

Conformance class 3: harmonized

 
 

  Overviews

thumbnail
large_thumbnail

  Provided by

  Share on social sites

         

  Views

  • INSPIRE
  • Simple
  • Full
  • XML
3fe7953d-daea-4a62-becd-6787f96f4405   Access to the portal Read here the full details and access to the data.

  Associated resources

Not available


  •  
  •  
  • API